TRUERESULT
Report
- Report Number
- 1052693-2015-01495
- Event Type
- Malfunction
- Date Received
- August 26, 2015
- Date of Event
- July 28, 2015
- Report Date
- October 8, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER AND TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD INACCURATE REFERENCE.
(B)(4). PRODUCT NOT YET EVALUATED.
CUSTOMER COMPLAINT OF HIGH BLOOD GLUCOSE RESULTS. CUSTOMER NORMALLY READS 89-120MG/DL BUT HAS BEEN CURRENTLY READING IN THE 400'S. CUSTOMER STATES HE FEELS FINE. DENIES ANY SYMPTOMS OF HIGH BLOOD SUGAR, DENIES NEED FOR MEDICAL ATTENTION. STRIP EXPIRATION DATE:09/26/2017. STRIP OPEN DATE: (B)(6) 2015. STRIP STORAGE:YES, STORED CORRECTLY. BACK TO BACK BLOOD GLUCOSE TESTING PERFORMED WITH RESULTS: BLOOD TEST 1: 130MG/DL FASTING:YES. BLOOD TEST 2: 120MG/DL FASTING:YES. REVIEWED METER MEMORY: 1: 332MG/DL (B)(6) 2015 07:27:00 AM FASTING:YES. 2: 365MG/DL (B)(6) 2015 07:44:00 PM FASTING:NO. 3: 220MG/DL (B)(6) 2015 07:58:00 AM FASTING:YES. 4: 361MG/DL (B)(6) 2015 07:48:00 AM FASTING:YES. 5 :228MG/DL (B)(6) 2015 05:04:00 PM FASTING:YES. CUSTOMERS CONCERN:365MG/DL. NO ADVERSE EVENT REPORTED.
CUSTOMER COMPLAINT OF HIGH BLOOD GLUCOSE RESULTS. CUSTOMER NORMALLY READS 89-120MG/DL BUT HAS BEEN CURRENTLY READING IN THE 400'S. CUSTOMER STATES HE FEELS FINE. DENIES ANY SYMPTOMS OF HIGH BLOOD SUGAR, DENIES NEED FOR MEDICAL ATTENTION. STRIP EXPIRATION DATE: 09/26/2017. STRIP OPEN DATE: (B)(6) 2015. STRIP STORAGE:YES, STORED CORRECTLY. BACK TO BACK BLOOD GLUCOSE TESTING PERFORMED WITH RESULTS: BLOOD TEST 1: 130MG/DL FASTING:YES. BLOOD TEST 2: 120MG/DL FASTING:YES. REVIEWED METER MEMORY: 332MG/DL (B)(6) 2015 07:27:00 AM FASTING:YES; 365MG/DL (B)(6) 2015 07:44:00 PM FASTING:NO; 220MG/DL (B)(6) 2015 07:58:00 AM FASTING:YES; 361MG/DL (B)(6) 2015 07:48:00 AM FASTING:YES; :228MG/DL (B)(6) 2015 05:04:00 PM FASTING:YES; CUSTOMERS CONCERN: 365MG/DL. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 564826 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PR2111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |