HYPERFORM BALLOON
Report
- Report Number
- 2029214-2015-00929
- Event Type
- Death
- Date Received
- August 25, 2015
- Date of Event
- July 22, 2015
- Report Date
- July 29, 2015
- Manufacturer
- COVIDIEN (IRVINE)
- Product Code
- MJN
- PMA / PMN Number
- K091458
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
THE CATHETER WAS RETURNED FOR EVALUATION WITH A GUIDEWIRE. THERE WAS A SIGNIFICANT AMOUNT OF BLOOD NOTED THROUGHOUT THE DEVICE AND THE INFLATION HOLES WERE OCCLUDED WITH DRIED BLOOD. THE CATHETER¿S INNER LINER AND COIL WERE FOUND TO BE STRETCHED WITH AN ACCORDION APPEARANCE AT THE DISTAL SECTION. THE BALLOON SUB-ASSEMBLY WAS ALSO FOUND TO BE ELONGATED WITH A NECKING APPEARANCE AT THE DISTAL SECTION AND AT PROXIMAL MARKER BAND SECTION. DUE TO THE CATHETER¿S DAMAGED CONDITION IT COULD NOT BE TESTED. NO OTHER ANOMALIES WERE OBSERVED. BASED ON THE ANALYSIS, THE CAUSE OF THE NON-DEFLATION COULD NOT BE DETERMINED; HOWEVER, IT IS POSSIBLE THAT BLOOD MAY HAVE ENTERED INTO THE BALLOON¿S SUB-ASSEMBLY DURING THE PROCEDURE OCCLUDING THE INFLATION HOLES CONTRIBUTING TO THE REPORTED EVENT. PER THE OCCLUSION BALLOON SYSTEM IFU: " DO NOT WITHDRAW THE GUIDEWIRE TO DEFLATE THE BALLOON; INSTEAD PULL NEGATIVE ON THE SYRINGE TO DEFLATE. THE BALLOON SHOULD ALWAYS BE DEFLATED WITH THE GUIDEWIRE REMAINING BEYOND THE CATHETER TIP IN ORDER TO MINIMIZE ASPIRATION OF BLOOD INTO THE LUMEN." IN ADDITION, AFTER MULTIPLE INFLATIONS, THE CATHETER AND GUIDEWIRE SHOULD BE REMOVED AND INSPECTED. VERIFY THE PERFORMANCE OF THE BALLOON AND INSPECT IT FOR THE PRESENCE OF CLOTS AT THE TIP OR INFLATION HOLES. THE LOT HISTORY RECORD OF THE REPORTED LOT NUMBER HAS BEEN REVIEWED AND NO QUALITY ISSUES WERE NOTED. REFERENCE (B)(4) FOLLOW UP 1.
THE LOT HISTORY RECORD OF THE REPORTED LOT NUMBER HAVE BEEN REVIEWED AND NO QUALITY ISSUES WERE NOTED.THE DEVICE HAS NOT BEEN RECEIVED. ONCE THE DEVICE IS RECEIVED A SUPPLEMENTAL REPORT WILL BE RESUBMITTED WITH THE ANALYSIS FINDINGS.(B)(4)
MEDTRONIC (COVIDIEN) RECEIVED INFORMATION THAT DURING THE TREATMENT OF A BASILAR ARTERY ANEURYSM THE HYPERFORM BALLOON FAILED TO DEFLATE DURING THE PROCEDURE. THE PHYSICIAN WAS INITIALLY ABLE TO DEFLATE AND INFLATE THE OCCLUSION BALLOON. AFTER READJUSTMENT OF THE MICROCATHETER POSITION THE FOLLOW UP RADIOGRAPHY REVEALED THAT THE ANEURYSM WAS BLEEDING. THE PHYSICIAN IMMEDIATELY INFLATED THE BALLOON TO OCCLUDE THE VESSEL AND QUICKLY FILL THE ANEURYSM WITH A COIL. HOWEVER UPON THE DETACHMENT OF THIS COIL, THE PHYSICIAN ATTEMPTED TO DEFLATE THE BALLOON BUT THE BALLOON DID NOT RESPOND TO THIS AND OTHER ATTEMPTS TO DEFLATE THE BALLOON. THE PHYSICIAN ULTIMATELY HAD TO PULL ON THE OCCLUSION BALLOON CATHETER TO REMOVE IT. AFTER REMOVAL OF THE BALLOON CATHETER THE FOLLOW UP RADIOGRAPHY REVEALED BLEEDING. THEREFORE, THE PHYSICIAN USED A NEW BALLOON SYSTEM TO CONTROL THE BLEEDING. THE PROCEDURE WAS THEN COMPLETED BUT THE PATIENT STILL WAS UNABLE TO BREATH SPONTANEOUS. THE PATIENT DID NOT HAVE IMPROVEMENT FROM BASELINE STATUS. PRIOR TO THE TREATMENT, THE PATIENT SUDDENLY BECAME UNCONSCIOUSNESS WITH NO CLEAR TRIGGER/CAUSE, WITH HYPERSPASMIA AND INCONTINENCE. THE PATIENT WAS TAKEN TO THE LOCAL HOSPITAL AND RECEIVED: CRANIAL COMPUTED TOMOGRAPHY (CT) - SUBARACHNOID HEMORRHAGE. THE PATIENT WAS INTUBATED, DEHYDRATION, AND PROVIDED WITH TREATMENT FOR PREVENTION OF CEREBRAL VASOSPASM. THE PATIENT WAS THEN TRANSFERRED TO (B)(6) FOR FURTHER TREATMENT. THE PATIENT WAS IN A COMA AND PRESENTED WITH BILATERAL DILATED PUPILS, WITHIN NO SPONTANEOUS BREATHING, LIMB MUSCLE STRENGTH WAS LEVEL FOUR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 562779 | HYPERFORM BALLOON | CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY | MJN | COVIDIEN (IRVINE) | 104-4420 | 9980598 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Death| R |