FDA Adverse Event Other Summary report: N

DRAGER DVE CEILING SERVICE UNIT

MDR report key: 50245 · Received November 5, 1996

Report

Report Number
2027060-1996-09001
Event Type
Other
Date Received
November 5, 1996
Date of Event
October 21, 1996
Report Date
October 30, 1996
Manufacturer
DRAGERWERK AG
Product Code
FQP
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING A PROCEDURE, THE PENDANT HEAD OF CEILING SERVICE UNIT SEPARATED FROM THE BEARING JOINT. TECHNICIAN AND HOSPITAL STAFF SECURED THE UNIT BY RESTING IT ON A CART SO THAT THERE WAS NO TENSION ON ANY OF THE LINES. THE PROCEDURE WAS COMPLETED WITH NO FURTHER DELAYS, COMPLICATIONS OR INTERRUPTION OF SERVICES. UPON INSPECTION, IT WAS FOUND THAT THE (4) M6X55MM BOLTS AND LOCK WASHERS WERE IN PLACE, BUT HAD BEEN UNSCREWED FROM PENDANT HEAD ROTATION BEARINGS. THE THREADED HOLES SHOW NO SIGN OF WEAR OR DAMAGE. THE LAST THREAD OF (1) BOLT WAS PULLED OFF IN THE SEPARATION. TOTAL SEPARATION WAS 4" AND OCCURRED AT LOCATION B. ALL NECESSARY REPAIRS WERE COMPLETED, AND INSPECTIONS WERE MADE OF ALL OTHER UNITS IN THE HOSPITAL. UPON INSPECTION, IT WAS FOUND THAT UNITS HAD HAD ADDITIONAL PORTS INSTALLED WITHOUT CO'S SUPERVISION OR SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRAGER DVE CEILING SERVICE UNIT CEILING SERVICE UNIT FQP DRAGERWERK AG DVE 4002-700MM/600MM NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other