FDA Adverse Event
Malfunction
Summary report: N
LTV
MDR report key: 5021816
·
Received August 21, 2015
Report
- Report Number
- 2031702-2015-00177
- Event Type
- Malfunction
- Date Received
- August 21, 2015
- Date of Event
- July 12, 2015
- Report Date
- August 20, 2015
- Manufacturer
- CAREFUSION 203, INC
- Product Code
- CBK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE CENTER REPLACED THE VENTILATOR'S FLOW VALVE TO CORRECT THE REPORTED PROBLEM. A VISUAL INSPECTION REVEALED THAT THE FLOW VALVE STEPPER MOTOR BAND WAS BROKEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE VENTILATOR HAD LOW PIP WITH AN AUDIBLE ALARM WHILE ON THE PATIENT. THE PIP WAS APPARENTLY READING FROM 15 TO 3 CMH2O AND WAS CHANGING WITH EACH BREATH BY AS MUCH AS 12 CMH2O. IT WAS ALSO REPORTED THAT THE VENTILATOR WAS HOT TO THE TOUCH. WHEN DOING A PROPER OPERATION CHECKOUT AFTER REMOVING THE VENTILATOR FROM THE PATIENT, IT WAS NOTICED THAT THE VTE WAS LESS THAN 50% OF THE DELIVERED TIDAL VOLUME. NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 554663 | LTV | VENTILATOR, CONTINUOUS | CBK | CAREFUSION 203, INC | LTV 900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |