FDA Adverse Event Malfunction Summary report: N

STAT COMFORT THIN LANCET - PULL OFF

MDR report key: 5021552 · Received August 13, 2015

Report

Report Number
1058955-2015-00018
Event Type
Malfunction
Date Received
August 13, 2015
Date of Event
March 28, 2014
Report Date
August 10, 2015
Manufacturer
STAT MEDICAL DEVICES, INC.
Product Code
FMK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

CUSTOMER COMPLAINED THAT IN ONE ARRIVAL OF LANCET 33G, BATCH 00314A CUSTOMER FOUND TWO OUTER OF LANCET 28G BATCH 00315A (CATALOG SM-SCR-100) BUT INSIDE THERE WERE BOXES OF BATCH 00314A. MOREOVER ONE BOX OF BATCH 00315A WAS FOUND. IMPROPER HANDLING CAUSED LOT CO-MINGLING AT MANUFACTURING FACILITY. LOT CO-MINGLING RESULTED FROM IMPROPER WORKPLACE CLEARANCE DURING THE PACKAGING OPERATIONS OF LOT 00314A, PRIOR TO INITIATION OF PACKAGING 00315A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
535563 STAT COMFORT THIN LANCET - PULL OFF TWIST OFF LANCET FMK STAT MEDICAL DEVICES, INC. SCU-100 00314A

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other