STAT COMFORT THIN LANCET - PULL OFF
MDR report key: 5021552
·
Received August 13, 2015
Report
- Report Number
- 1058955-2015-00018
- Event Type
- Malfunction
- Date Received
- August 13, 2015
- Date of Event
- March 28, 2014
- Report Date
- August 10, 2015
- Manufacturer
- STAT MEDICAL DEVICES, INC.
- Product Code
- FMK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
CUSTOMER COMPLAINED THAT IN ONE ARRIVAL OF LANCET 33G, BATCH 00314A CUSTOMER FOUND TWO OUTER OF LANCET 28G BATCH 00315A (CATALOG SM-SCR-100) BUT INSIDE THERE WERE BOXES OF BATCH 00314A. MOREOVER ONE BOX OF BATCH 00315A WAS FOUND. IMPROPER HANDLING CAUSED LOT CO-MINGLING AT MANUFACTURING FACILITY. LOT CO-MINGLING RESULTED FROM IMPROPER WORKPLACE CLEARANCE DURING THE PACKAGING OPERATIONS OF LOT 00314A, PRIOR TO INITIATION OF PACKAGING 00315A.