FDA Adverse Event
Injury
Summary report: N
9614385-2003-00187
MDR report key: 502145
·
Received December 12, 2003
Report
- Report Number
- 9614385-2003-00187
- Event Type
- Injury
- Date Received
- December 12, 2003
- Date of Event
- October 9, 2003
- Product Code
- FWM
- Report Source
- Manufacturer report
Narratives
Description of Event or Problem · 1
DEFLATION OF RIGHT SALINE BREAST IMPLANT FOLLOWED BY REPLACEMENT WITH ANOTHER BRAND. UNKNOWN IF INITIAL USE. ORIGINAL SURGERY WAS PERFORMED IN 2000 USING HUTCHISON HSH 450 SALINE-FILL IMPLANTS, FILLED TO A VOLUME OF 550CC WHICH EXCEEDS THE RECOMMENDED FILL VOLUME LIMIT OF 450-480CC BY 70CC. PT UNDERWENT UNILATERAL REPLACEMENT SURGERY IN 2003, WITH RIGHT IMPLANT BEING REPLACED WITH MCGHAN STYLE 68 SMOOTH ROUND SALINE-FILL BREAST IMPLANT, 550CC, FILLED TO 550CC.