VERSACARE BED
Report
- Report Number
- 1824206-2015-00832
- Event Type
- Malfunction
- Date Received
- August 21, 2015
- Date of Event
- August 3, 2015
- Report Date
- August 3, 2015
- Manufacturer
- HILL-ROM BATESVILLE
- Product Code
- FNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICIAN FOUND THE EXTERNAL ALARM WAS FAULTY FOR THIS DEVICE. PER THE HILL-ROM USER MANUAL, WARNING: THE BED EXIT ALARM SYSTEM IS NOT INTENDED AS A SUBSTITUTE FOR GOOD NURSING PRACTICES. THE BED EXIT ALARM SYSTEM MUST BE USED IN CONJUNCTION WITH A SOUND RISK ASSESSMENT AND PROTOCOL. PER THE HILL-ROM USER MANUAL, IF THE BED EXIT ALARM DOES NOT ARM AND ALL THREE MODE INDICATORS ARE FLASHING, REMOVE THE PATIENT AND ZERO THE BED EXIT SYSTEM. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE EXTERNAL ALARM TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED ALARM HAD NO SOUND. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 556367 | VERSACARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNK | HILL-ROM BATESVILLE | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |