VERSACARE BED
Report
- Report Number
- 1824206-2015-00831
- Event Type
- Malfunction
- Date Received
- August 21, 2015
- Date of Event
- August 3, 2015
- Report Date
- August 3, 2015
- Manufacturer
- HILL-ROM BATESVILLE
- Product Code
- FNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICAL SUPPORT FOUND THE BRAKE OUT CABLE HAD BROKEN PINS CAUSING THE NURSE CALL NOT TO FUNCTION CORRECTLY. PER THE HILL-ROM USER MANUAL, WARNING: A COMMUNICATION CABLE MUST BE USED FOR BEDS THAT HAVE NURSE CALL. FAILURE TO DO SO COULD CAUSE PATIENT INJURY. FOR BEDS THAT HAVE NURSE CALL SYSTEMS, A COMMUNICATION CABLE MUST BE CONNECTED BETWEEN THE BED AND FACILITY COMMUNICATION SYSTEM. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2008-2010, 2012, AND 2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE ACCOUNT REPLACED THE CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE NURSE CALL NOT WORKING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 556090 | VERSACARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNK | HILL-ROM BATESVILLE | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |