HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Report
- Report Number
- 3004365956-2015-00264
- Event Type
- Malfunction
- Date Received
- August 20, 2015
- Date of Event
- July 29, 2015
- Report Date
- August 7, 2015
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- CAF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). NO VISUAL OR DIMENSIONAL INSPECTION CAN BE PERFORMED SINCE THE DEVICE SAMPLE IS NOT AVAILABLE FOR EVALUATION. NO FUNCTIONAL INSPECTION CAN BE PERFORMED SINCE THE DEVICE SAMPLE IS NOT AVAILABLE FOR EVALUATION. FAILURE MODE COULD NOT BE DUPLICATED SINCE IT IS UNKNOWN THE REASON THE ADAPTOR IS NOT FITTING CORRECTLY TO THE OXYGEN FLOW METER, HOWEVER FUNCTIONAL TEST WAS PERFORMED ON 30 SUBASSEMBLY (P/N: 12161) & NO ISSUES OR DISCREPANCIES WERE FOUND. THE TEST CONSISTED OF 30 PIECES FROM THE PROCESS, VISUAL INSPECTION AND OXYGEN FLOWMETER WAS USED WITH THE SAMPLES AND IT WAS FOUND THEY PERFORMED ACCORDING TO SPECIFICATION. A DHR (DEVICE HISTORY RECORD) REVIEW COULD NOT BE CONDUCTED SINCE THE LOT NUMBER WAS NOT PROVIDED. BASED ON SIMILAR COMPLAINTS A CAPA FILE# (B)(4) WAS OPENED TO FURTHER INVESTIGATE THIS ISSUE. NO CONCLUSION CAN BE ESTABLISHED AT THIS TIME BASED ON THE LACK OF DEVICE SAMPLE. IF THE DEVICE SAMPLE BECOMES AVAILABLE THIS COMPLAINT WILL BE REOPENED.
(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL EXAM WAS PERFORMED AND IT WAS OBSERVED THAT THE INTERNAL THREADS OF THE ADAPTOR WERE DAMAGED. OXYGEN ENTRAINMENT TESTING COULD NOT BE PERFORMED AS THE ADAPTOR COULD NOT BE CONNECTED CORRECTLY ON THE OXYGEN SUPPLY DUE THE DAMAGE ON THE INTERNAL THREAD. BASED ON THE VISUAL EXAM, THE REPORTED COMPLAINT WAS CONFIRMED. THE ROOT CAUSE FOR THE ISSUE IS THE POSITIONING OF THE THREAD LEAD AND THE SOFTNESS OF THE NEW RESIN USED FOR THE SNAP ADAPTOR. A CAPA WAS OPENED TO ADDRESS THIS ISSUE.
THE CUSTOMER ALLEGES THAT THE ADAPTOR DIDN'T FIT WITH THE OXYGEN FLOWMETER. NO PATIENT INJURY REPORTED.
THE CUSTOMER ALLEGES THAT THE ADAPTOR DIDN'T FIT WITH THE OXYGEN FLOWMETER. NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 551724 | HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE | ADAPTOR | CAF | TELEFLEX MEDICAL | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |