HUT EXT DR FINAL ASSY-REVERSE
Report
- Report Number
- 1518293-2015-00079
- Event Type
- Malfunction
- Date Received
- August 20, 2015
- Date of Event
- July 31, 2015
- Report Date
- July 31, 2015
- Manufacturer
- LIEBEL FLARSHEIM
- Product Code
- IXR
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
TECH SUPPORT SPOKE WITH THE CUSTOMER WHO SAID THE OPERATORS COULD NOT GET THE INFIMED I5 TO OPERATE AND HAD TO MOVE THE PATIENT TO A C-ARM TO COMPLETE THE CASE. THE INFIMED WAS DISPLAYING THE MESSAGE - GIM COMMUNICATIONS FAILURE AND THE KEYBOARD/MOUSE WAS LOCKED UP. TECH SERVICE HAD OPERATOR CYCLE THE POWER TO THE GIM BOX BY PHYSICALLY UNPLUGGING/PLUGGING IT BACK IN AND THEN CYCLING POWER ON THE I5 TOWER. ON POWER UP, EVERYTHING WAS NOW WITHOUT ISSUE AND OPERATOR WAS ABLE TO OPEN UP A PATIENT AND PERFORM X-RAY. ALL OF THIS INDICATES THAT THE ORIGINAL ISSUE WAS MOST LIKELY A GIM BOX COMMUNICATIONS ISSUE AND THIS ISSUE WAS CORRECTED BY CYCLING POWER AT THE GIM. THERE ARE CURRENTLY NO OTHER REPORTED ISSUES WITH THIS SYSTEM.
CUSTOMER REPORTS THE FLUORO FAILED DURING A STENT PLACEMENT WITH A GIM COMMUNICATION ERROR. THE PATIENT WAS TRANSFERRED TO ANOTHER ROOM AND THE PROCEDURE WAS COMPLETED WITH A C-ARM. NO REPORTED INJURY.