FDA Adverse Event Malfunction Summary report: N

QWIKLET ULTIMATE LANCING DEVICE

MDR report key: 5018701 · Received August 12, 2015

Report

Report Number
1058955-2015-00011
Event Type
Malfunction
Date Received
August 12, 2015
Date of Event
June 16, 2011
Report Date
August 5, 2015
Manufacturer
STAT MEDICAL DEVICES, INC.
Product Code
FMK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE ORIGINAL (B)(4), FOR THIS PRODUCT RETURN WAS CLOSED ON 08/24/2011. THE CUSTOMER THEN RETURNED 12 DEVICES CLAIMING THEY WERE THE ACTUAL FAILURES. THE COMPLAINT WAS REOPENED AND CLOSED AGAIN 09/14/2011. THIS NEW (B)(4) WAS OPENED AFTER CONSULTATION WITH THE (B)(4) ON THE RE-ASSESSMENT OF THE MDR SUBMISSION FOR THIS RETURNED PRODUCT.

Description of Event or Problem · 1

CUSTOMER HAS INDICATED THEY HAVE RECEIVED A DOZEN COMPLAINTS FROM THEIR CUSTOMERS THAT THE DEVICES ARE BREAKING AND DO NOT PENETRATE THE SKIN. THE CUSTOMER INDICATED THE SAMPLES COULD NOT BE RETURNED. TWENTY ULTIMATE LANCING DEVICE SAMPLES FROM VARIOUS LOTS WERE TAKEN FROM STAT INVENTORY AND TESTED BY COCKING AND FIRING THE DEVICE AT EACH OF THE EIGHT (8) DEPTH SETTINGS INTO 0.09 INCH THICK STACKS OF PAPER. THE NUMBER OF PENETRATED PAGES WAS COUNTED; ALL 20 SAMPLES COCKED SUCCESSFULLY 100% OF THE TRIALS AND MET THE PRODUCTION SPECIFICATION FOR PAPER PENETRATION. ON 09/14/2011, THE CUSTOMER RETURNED TWELVE (12) ULTIMATE LANCING DEVICES WITH NO OTHER INFORMATION. EACH DEVICE WAS JAMMED STUCK IN THE COCKED POSITION. AFTER MANUALLY UNJAMMING THE COCKING MECHANISMS, ALL RETURNED SAMPLES PASSED THE COCKING PENETRATION TESTS DESCRIBED EARLIER. THE JAMMED CONDITION COULD NOT BE REPRODUCED, EVEN WITH MISUSE MANIPULATION BY INSERTING THE LANCET WITH EXTREME FORCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
531119 QWIKLET ULTIMATE LANCING DEVICE LANCING DEVICE FMK STAT MEDICAL DEVICES, INC. QLU-L02 UNK

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other