FDA Adverse Event Injury Summary report: N

LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI)

MDR report key: 5017334 · Received August 18, 2015

Report

Report Number
2023950-2015-00129
Event Type
Injury
Date Received
August 18, 2015
Date of Event
July 16, 2015
Report Date
August 18, 2015
Manufacturer
ZEST ANCHORS, INC.
Product Code
DZE
PMA / PMN Number
K120198
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

USER DOCUMENTATION ((B)(4)) SPECIFIES THAT IMMEDIATE LOADING IS SUITABLE ONLY IF SUFFICIENT PRIMARY STABILITY OF THE IMPLANT IS ACHIEVED AT THE TIME OF PLACEMENT. ALSO, THE RECOMMENDED IMPLANT INSERTION TORQUE AND IMPLANT ATTACHMENT TORQUE IS 30N-CM. IF THE IMPLANT PLACEMENT TORQUE IS LESS THAN 30 N-CM, THEN THE ATTACHMENTS SHOULD ONLY BE HAND TIGHTENED. THE IMPLANT INSERTION TORQUE SHOULD NOT EXCEED 70 N-CM; IF 70 N-CM IS REACHED PRIOR TO FULL SEATING, THE IMPLANT SHOULD BE REMOVED AND THE OSTEOTOMY SHOULD BE ENLARGED. CLINICIAN STATED THAT A LODI IMPLANT FAILED TO OSSEOINTEGRATE. PT HAD AN INFECTION AND HAS MODERATE DENSITY BONE. IMPLANT WAS IMMEDIATELY LOADED. THE CLINICIAN DID NOT PROVIDE THE FOLLOWING INFO: AMOUNT OF TORQUE USED ON ABUTMENT AND IMPLANT; WHETHER PRIMARY STABILITY WAS ACHIEVED. FAILURE TO OSSEOINTEGRATE IS A WELL-DOCUMENTED INHERENT RISK OF DENTAL IMPLANTS. IN THE MAJORITY OF CASES WHERE AN IMPLANT FAILS TO INTEGRATE WITH THE BONE AND IS REJECTED BY THE BODY, THE CAUSE IS UNKNOWN. THE RECORDS MANAGEMENT DATABASE OF THE IMPLANT INDICATES THAT LL IMPLANTS MANUFACTURED IN THIS SPECIFIC LOT MET THE SPECIFICATIONS AND WERE APPROVED FOR PRODUCT RELEASE. ANY ISSUES WERE SUCCESSFULLY RESOLVED PRIOR TO THE RELEASE OF THE IMPLANTS. NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

LODI IMPLANT FAILED TO OSSEOINTEGRATE. PT ALSO HAD AN INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
543815 LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI) ENDOSSEOUS DENTAL IMPLANT DZE ZEST ANCHORS, INC. I0RQRQ

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention