TI PARALLEL CONNECTOR 6.0MM/6.0MM
Report
- Report Number
- 2530088-2015-10554
- Event Type
- Injury
- Date Received
- August 19, 2015
- Report Date
- August 3, 2015
- Manufacturer
- SYNTHES BRANDYWINE
- Product Code
- NKB
- PMA / PMN Number
- PK082572
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NURSE
Narratives
(B)(4): SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ONGOING; NO CONCLUSION COULD BE DRAWN AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM.A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED.DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PATIENT¿S WEIGHT IS UNKNOWN BUT REPORTED AS NORMAL (B)(6). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
A PRODUCT INVESTIGATION WAS PERFORMED ¿ FIVE 6.0MM PANGEA POLYAXIAL SCREWS (PART# 04.620.645, LOT# 7059241, 7177835, 5883928, 7182324, 7177838), FIVE LOCKING CAPS (PART# 04.620.000, LOT# 7212831 QTY 3, 7183995 QTY 2), ONE TRANSCONNECTOR (PART# 04.620.204, LOT# 6690662), TWO PARALLEL CONNECTORS (PART# 498.160, LOT# 6835464, QTY 2) ONE EXTENSION CONNECTOR (PART# 498.165, LOT# 7695223) AND TWO TITANIUM RODS (PART# 498.106 LOT #: 6430550, PART# 498.105 LOT #: 0784868) WERE RETURNED FOR THE SCREWS NOT BEING WELL IMPLANTED AND THE PATIENT'S BONE NOT GROWING AROUND THEM. THIS COULD BE A RESULT OF POOR BONE QUALITY, THE LOCKING CAPS NOT BEING TIGHTENED TO THE APPROPRIATE TORQUE, OR A REACTION TO THE TAN ALLOY OF THE SCREWS. REPLICATION OF THE COMPLAINT CONDITION IS NOT APPLICABLE AND THE COMPLAINT IS CONFIRMED. THE RODS WERE RETURNED CHIPPED AND NICKED AND ALSO HAD THE ANODIZATION WORN OFF IN SPOTS. THE TRANSCONNECTOR WAS RETURNED WITH THE SCREW HOLDING THE TWO HOOKS TOGETHER MISSING. NO OTHER DAMAGE VISIBLE ON THE COMPONENTS. THE TWO PARALLEL CONNECTORS WERE RETURNED WITH ALL OF THE SET SCREWS AND MINIMAL DAMAGE. THE LOCKING CAPS CONTAINED MINIMAL DAMAGE BUT SHOWED WEAR ON THE RIDGED SURFACE THAT INTERFACES WITH THE RODS. THE STRAIGHT CONNECTOR HAD MINIMAL DAMAGE AND ALL SET SCREWS. IT IS UNLIKELY THE RETURNED LOCKING CAPS, RODS, AND CONNECTORS CONTRIBUTED TO THIS COMPLAINT AS THEY DO NOT DIRECTLY AFFECT THE SCREW PURCHASE. THE DETAILS REGARDING THE TECHNIQUE USED AND THE BONE QUALITY OF THE PATIENT ARE UNKNOWN AND SO THE ROOT CAUSE CANNOT BE DETERMINED. DHR REVIEW ¿ MANUFACTURING DATE: 7DEC2011. PART 498.160, LOT# 6835464 DID NOT CONTAIN ANY NCRS OR ANY ANOMALIES. ALL LOT PIECES WERE ACCEPTED/CONFORMING. A DHR REVIEW FOR THE DEVICE COMPONENTS WERE REVIEWED: BOTH LOTS DID NOT CONTAIN ANY NCRS OR ANY ANOMALIES. ALL PIECES WERE ACCEPTED/CONFORMING. FOR PART E2062 SOME COMPONENTS WERE SCRAPPED AS THEY WERE USED FOR INSPECTION SAMPLING. A DHR REVIEW FOR THE RAW MATERIAL WAS ALSO PERFORMED: BOTH LOTS DID NOT CONTAIN ANY NCRS OR ANOMALIES AND ALL WERE ACCEPTED/CONFORMING AND RELEASED. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES (B)(04) REPORTED THE FOLLOWING EVENT: IT WAS REPORTED A REMOVAL SURGERY WAS PERFORMED. THIS WAS THE PATIENT'S THIRD SURGERY; THE FIRST SURGERY WAS ON (B)(6) 2013 AND THE SECOND SURGERY WAS ON (B)(6) 2013. THE FIRST SURGERY ENTAILED AN ILIAC-L1 FUSION UTILIZING THE PANGEA SYSTEM. THE SECOND SURGERY WAS PERFORMED DUE TO THE FACT THAT TWO OF THE INITIAL SCREWS WERE PULLING OUT OF THE BONE DUE TO THE PATIENT'S POOR BONE QUALITY AND ENTAILED T11-T8 EXTENSION FROM THE FIRST SURGERY ALSO USING THE PANGEA SYSTEM, WITH DOMINOS. THE TWO (2) SCREWS THAT WERE PULLING OUT WERE LEFT IN THE PATIENT. APPROXIMATELY TWO YEARS LATER, THE PATIENT COMPLAINED OF BACK PAIN AND THE SURGEON REALIZED THAT THE SCREWS INSERTED FROM T11-T8 WERE NOT WELL IMPLANTED INTO THE BONE. UPON MRI, THE SURGEON DISCOVERED THAT AROUND THE SCREWS A VOID WAS BEING CREATED AND THE BONE WAS NOT GROWING AROUND THE SCREW, HENCE NO FUSION WAS OCCURRING. CONSEQUENTLY ON (B)(6) 2015, THE SURGEON PERFORMED A REMOVAL SURGERY WHEN SCREWS FROM T11-T8 WERE REMOVED ALONG WITH DOMINOS, CROSSLINKS, SCREWS AND CAPS. ALL ORIGINALLY IMPLANTED SCREWS WERE LEFT IN THE PATIENT AS FUSION HAD OCCURRED AT THE SITE WHERE THE ORIGINAL SCREWS HAD PULLED OUT. THE SURGEON KEPT TWO SCREWS IN ORDER TO PERFORM ALLERGY TESTS. IT WAS ALSO REPORTED THAT THE SCREWS WERE REMOVED VERY EASILY. THE PATIENT OUTCOME IS UNKNOWN. THIS IS REPORT 13 OF 16 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 546811 | TI PARALLEL CONNECTOR 6.0MM/6.0MM | ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD | NKB | SYNTHES BRANDYWINE | 6835464 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |