FDA Adverse Event Malfunction Summary report: N

RECONDITIONED TOTALCARE BED

MDR report key: 5016128 · Received August 19, 2015

Report

Report Number
1824206-2015-00821
Event Type
Malfunction
Date Received
August 19, 2015
Date of Event
July 24, 2015
Report Date
July 24, 2015
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE BIOMEDICAL ENGINEER FOUND THE BRAKE CASTER WAS WORN. PER THE HILL-ROM SERVICE MANUAL THE TOTALCARE BED SYSTEM REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM A SEMI-ANNUAL PREVENTATIVE MAINTENANCE. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE BRAKE CASTERS TO DETERMINE IF THE BED MOVES WHEN YOU ACTIVATE THE BRAKE, REPLACE OR ADJUST WHEN NECESSARY. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE BIOMEDICAL ENGINEER REPLACED THE BRAKE CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT STATING, "THE CASTER ROLLS AND SWIVELS WHEN LOCKED INTO BRAKE. PEDAL IS WORKING, CASTER WILL NOT LOCK." THERE WAS NO PATIENT OR USER INJURY RELATED TO THIS COMPLAINT. THIS COMPLAINT HAS BEEN DOCUMENTED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
548835 RECONDITIONED TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC.

Patients

Seq Age Sex Outcome Treatment
1