FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 5014797 · Received August 13, 2015

Report

Report Number
1052693-2015-01404
Event Type
Malfunction
Date Received
August 13, 2015
Date of Event
July 16, 2015
Report Date
December 4, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT UNDER EVAL.

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION IMPROPER STORAGE OF TEST STRIPS AS INDICATED BY CUSTOMER. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF "LO" BLOOD GLUCOSE TEST RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 145 MG/DL. TESTING PERFORMED ONCE DAILY IN THE MORNING FASTING. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THE TIME OF THE CALL ON (B)(6) 2015 AS A RESULT OF THE METER'S READINGS. CURRENTLY NOT TAKING MEDICATION TO MANAGE DIABETES. VERIFIED STORAGE OF PRODUCT IS NOT WITHIN INSTRUCTED SPECIFICATION SINCE THEY ARE RETAINED IN THE KITCHEN. TEST STRIP LOT MFR'S EXPIRATION DATE IS 09/17/2016 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: LO (B)(6) 2015 09:04AM; LO (B)(6) 2015 09:02AM; 157MG/DL (B)(6) 2015 08:49AM; 155MG/DL (B)(6) 2015 08:48AM; 142MG/DL (B)(6) 2015 09:25AM. ADVERSE EVENT NOT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF "LO" BLOOD GLUCOSE TEST RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 145 MG/DL. TESTING PERFORMED ONCE DAILY IN THE MORNING FASTING. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THE TIME OF THE CALL ON (B)(6) 2015 AS A RESULT OF THE METER'S READINGS. CURRENTLY NOT TAKING MEDICATION TO MANAGE DIABETES. VERIFIED STORAGE OF PRODUCT IS NOT WITHIN INSTRUCTED SPECIFICATION SINCE THEY ARE RETAINED IN THE KITCHEN. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/17/2016 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 1:LO (B)(6) 2015 09:04AM. 2:LO (B)(6) 2015 09:02AM. 3:157MG/DL (B)(6) 2015 08:49AM. 4:155MG/DL (B)(6) 2015 08:48AM. 5:142MG/DL (B)(6) 2015 09:25AM. ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
535437 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PP1580

Patients

Seq Age Sex Outcome Treatment
1 0 YR