TRUERESULT
Report
- Report Number
- 1052693-2015-01390
- Event Type
- Malfunction
- Date Received
- August 13, 2015
- Date of Event
- July 15, 2015
- Report Date
- December 22, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4) MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION STRIP ISSUE OR USER HAS HIGH GLUCOSE VALUE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.
(B)(4). RETURNED PRODUCT PENDING EVALUATION RESULTS.
CONSUMER COMPLAINT OF BLOOD RESULT READING OF "HI". CUSTOMER STATES THAT HE TESTED WITH HIS TRUERESULT METER ON HIS GRANDSON AND OBTAINED A "HI"-FASTING THREE TIMES. HIS GRANDSON IS NOT A DIABETIC. CUSTOMER STATES HIS GRANDSON FEELS WELL, BUT THEY SET UP AN APPOINTMENT TO SEE PHYSICIAN DUE TO THESE RESULTS. VERIFIED THE STRIPS EXPIRE 06/23/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED (B)(6) 2015. REVIEWED METER MEMORY: 1:HI (B)(6) 2015 11:40:00 AM FASTING:YES. 2:HI (B)(6) 2015 04:00:00 PM FASTING:YES. 3:HI (B)(6) 2015 04:00:00 PM FASTING:YES. CUSTOMER DISCLOSED THAT BLOOD TEST WERE PERFORMED USING AN ALTERNATE SITE, THE FOREARM.
CONSUMER COMPLAINT OF BLOOD RESULT READING OF "HI". CUSTOMER STATES THAT HE TESTED WITH HIS TRUE RESULT METER ON HIS GRANDSON AND OBTAINED A "HI"-FASTING THREE TIMES. HIS GRANDSON IS NOT A DIABETIC. CUSTOMER STATES HIS GRANDSON FEELS WELL, BUT THEY SET UP AN APPOINTMENT TO SEE PHYSICIAN DUE TO THESE RESULTS. VERIFIED THE STRIPS EXPIRE 06/23/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED (B)(6) 2015. REVIEWED METER MEMORY: HI, (B)(6) 2015 11:40:00 AM FASTING: YES; HI, (B)(6) 2015 04:00:00 PM FASTING: YES; HI, (B)(6) 2015 04:00:00 PM FASTING: YES. CUSTOMER DISCLOSED THAT BLOOD TEST WERE PERFORMED USING AN ALTERNATE SITE, THE FOREARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 532532 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PR2016 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |