FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 5014618 · Received August 13, 2015

Report

Report Number
1052693-2015-01390
Event Type
Malfunction
Date Received
August 13, 2015
Date of Event
July 15, 2015
Report Date
December 22, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION STRIP ISSUE OR USER HAS HIGH GLUCOSE VALUE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Additional Manufacturer Narrative · 1

(B)(4). RETURNED PRODUCT PENDING EVALUATION RESULTS.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF BLOOD RESULT READING OF "HI". CUSTOMER STATES THAT HE TESTED WITH HIS TRUERESULT METER ON HIS GRANDSON AND OBTAINED A "HI"-FASTING THREE TIMES. HIS GRANDSON IS NOT A DIABETIC. CUSTOMER STATES HIS GRANDSON FEELS WELL, BUT THEY SET UP AN APPOINTMENT TO SEE PHYSICIAN DUE TO THESE RESULTS. VERIFIED THE STRIPS EXPIRE 06/23/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED (B)(6) 2015. REVIEWED METER MEMORY: 1:HI (B)(6) 2015 11:40:00 AM FASTING:YES. 2:HI (B)(6) 2015 04:00:00 PM FASTING:YES. 3:HI (B)(6) 2015 04:00:00 PM FASTING:YES. CUSTOMER DISCLOSED THAT BLOOD TEST WERE PERFORMED USING AN ALTERNATE SITE, THE FOREARM.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF BLOOD RESULT READING OF "HI". CUSTOMER STATES THAT HE TESTED WITH HIS TRUE RESULT METER ON HIS GRANDSON AND OBTAINED A "HI"-FASTING THREE TIMES. HIS GRANDSON IS NOT A DIABETIC. CUSTOMER STATES HIS GRANDSON FEELS WELL, BUT THEY SET UP AN APPOINTMENT TO SEE PHYSICIAN DUE TO THESE RESULTS. VERIFIED THE STRIPS EXPIRE 06/23/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED (B)(6) 2015. REVIEWED METER MEMORY: HI, (B)(6) 2015 11:40:00 AM FASTING: YES; HI, (B)(6) 2015 04:00:00 PM FASTING: YES; HI, (B)(6) 2015 04:00:00 PM FASTING: YES. CUSTOMER DISCLOSED THAT BLOOD TEST WERE PERFORMED USING AN ALTERNATE SITE, THE FOREARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
532532 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PR2016

Patients

Seq Age Sex Outcome Treatment
1 0 YR