FEMORAL TRIAL SLAPHAMMER
Report
- Report Number
- 3005985723-2015-00144
- Event Type
- Malfunction
- Date Received
- August 19, 2015
- Date of Event
- July 22, 2015
- Report Date
- July 22, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- NPJ
- PMA / PMN Number
- K090763
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVENT REGARDING FRACTURE OF A FEMORAL TRIAL SLAPHAMMER WAS REPORTED. THE EVENT WAS NOT CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION COULD NOT BE PERFORMED AS THE SUBJECT DEVICE WAS NOT RETURNED. MEDICAL RECORDS RECEIVED AND EVALUATION: NO MEDICAL RECORDS WERE RECEIVED FOR REVIEW WITH A CLINICAL CONSULTANT. DEVICE HISTORY REVIEW INDICATED THE DEVICES ACCEPTED INTO FINAL STOCK FROM THE REPORTED LOT WERE FREE FROM DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAS BEEN NO OTHER EVENTS FOR THE LOT REFERENCED. CONCLUSION: THE EVENT COULD NOT BE CONFIRMED BUT BASED ON THE PROVIDED INFORMATION, THE ROOT CAUSE WAS LIKELY DUE TO USER ERROR AS THE FEMORAL TRIAL IS INTENDED TO BE IMPACTED, NOT THE REPORTED SLAPHAMMER. NO FURTHER INVESTIGATION IS POSSIBLE AT THIS TIME. IF THE DEVICES AND/OR ADDITIONAL INFORMATION ARE RECEIVED, THIS INVESTIGATION WILL BE REOPENED AND RE-EVALUATED.
AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS RETAINED BY THE HOSPITAL AND WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT MADE AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE IT WILL BE REPORTED IN A SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION. NOT RETURNED TO THE MANUFACTURER
DURING INSERTION OF THE FEMORAL TRIAL, DR. USED A MALLET TO STRIKE THE 111140 INSERTER/EXTRACTOR AND THE THREADED TIP BROKE OFF INSIDE THE TRIAL. NO ISSUES AROSE FROM THIS INCIDENT AS DR. WAS ABLE TO REMOVE THE TRIAL BY HAND. I ADVISED HIM THAT THE INSTRUMENT WAS NOT INTENDED FOR IMPACTION WITH A MALLET AS WE PROVIDE A SEPARATE IMPACTOR IN THE TRAY.
DURING INSERTION OF THE FEMORAL TRIAL, DR. USED A MALLET TO STRIKE THE 111140 INSERTER/EXTRACTOR AND THE THREADED TIP BROKE OFF INSIDE THE TRIAL. NO ISSUES AROSE FROM THIS INCIDENT AS DR. WAS ABLE TO REMOVE THE TRIAL BY HAND. I ADVISED HIM THAT THE INSTRUMENT WAS NOT INTENDED FOR IMPACTION WITH A MALLET AS WE PROVIDE A SEPARATE IMPACTOR IN THE TRAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 547949 | FEMORAL TRIAL SLAPHAMMER | INSTRUMENT | NPJ | MAKO SURGICAL CORP. | 06030813 | 06030813 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |