G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-75696
- Event Type
- Malfunction
- Date Received
- August 18, 2015
- Date of Event
- July 30, 2015
- Report Date
- July 30, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000019
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT SENSOR WAS NOT RETURNED TO DEXCOM. THE PEDIATRIC SHARE RECEIVER (PART NUMBER STK-PR-001/SERIAL NUMBER (B)(4)/LOT NUMBER 5201389), BEING USED WITH THE COMPLAINT SENSOR, WAS RETURNED ON (B)(6) 2015. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DIAGNOSTIC CHART CONFIRMED THE REPORTED EVENT OF INACCURACIES. A ROOT CAUSE COULD NOT BE DETERMINED.
(B)(4). IT WAS REPORTED THAT THE SENSOR WAS INSERTED INTO THE ARM. THE DEXCOM G4® PLATINUM (PEDIATRIC) CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: DO NOT INSERT THE SENSOR IN SITES OTHER THAN THE BELLY (ABDOMEN) OR UPPER BUTTOCKS. USE IN OTHER SITES MIGHT CAUSE SENSOR GLUCOSE READINGS TO BE INACCURATE AND COULD RESULT IN YOU MISSING SEVERE HYPOGLYCEMIA (LOW BLOOD GLUCOSE) OR HYPERGLYCEMIA (HIGH BLOOD GLUCOSE) EVENTS.
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE (BG) METER THAT OCCURRED ON (B)(6) 2015. SENSOR WAS INSERTED ON (B)(6) 2015 INTO THE ARM. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 545667 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-27 | 5199104 | 00386270000019 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |