GENTLEPOWER LUX CONTRA ANGLE 25LPR
Report
- Report Number
- 3003637274-2015-00021
- Event Type
- Injury
- Date Received
- August 18, 2015
- Date of Event
- July 14, 2015
- Report Date
- July 20, 2015
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EKX
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- DENTIST
Narratives
THE TEST RUN DURING THE ANALYSIS OF THE REPAIR DID SHOW THAT THE HEAD HEATED UP QUICKLY. THE ROOT CAUSE WAS THAT THE HEAD HAD A DENT WHICH CAUSED THE PUSH BUTTON TO STICK IN, RUBBING ON THE DRIVE ASSEMBLY CAUSING UNUSUAL FRICTION AND THEREFORE INCREASE OF TEMPERATURE. THE DENT SHOWS THAT THE HANDPIECE RECEIVED A STRONG HIT FROM OUTSIDE, E.G. DUE TO A DROP. THE TEST RUN DID ALSO SHOW THAT THE CURRENT CONSUMPTION WAS OUT OF SPECIFICATION. ROOT CAUSE WAS ON ONE HAND THE RUBBING PUSH BUTTON BUT ON THE OTHER HAND THE CONDITION OF THE BEARINGS. THEY HAVE BEEN FOUND GRITTY AFTER DISASSEMBLING THE HANDPIECE WHICH CAUSED IN ADDITION HIGHER FRICTION AND HENCE INCREASE OF TEMPERATURE. TO AVOID SUCH ISSUES THE USER INSTRUCTION CONTAINS ALREADY SEVERAL NOTES, WARNINGS AND REQUESTS HOW TO PREPARE THE HANDPIECE FOR EACH TREATMENT: WARNING: HAZARDS FOR THE CARE PROVIDER AND THE PATIENT. IN THE CASE OF DAMAGE, IRREGULAR RUNNING NOISE, EXCESSIVE VIBRATION, UNTYPICAL WARMING OR WHEN THE CUTTER OR GRINDER CANNOT BE HELD. DO NOT USE FURTHER AND NOTIFY SERVICE. CAUTION: BURNING HAZARD FROM HOT INSTRUMENT HEAD OR HOT INSTRUMENTS COVER. IF THE INSTRUMENT OVERHEATS, BURNS MAY ARISE IN THE ORAL AREA. NEVER CONTACT SOFT TISSUE WITH THE INSTRUMENT HEAD OR INSTRUMENT COVER THE FOLLOWING GUIDELINES MUST BE OBSERVED TO ENSURE SAVE USE OF THE ELECTRICALLY DRIVEN CONTRA-ANGLE HANDPIECES: THE SERVICE INSTRUCTIONS FOR CONTRA-ANGLE HANDPIECES MUST BE PRECISELY FOLLOWING WHEN USING KAVO SPRAY OR QUATTROCARE CARE SYSTEMS. BEFORE EACH USE, THE CONTRA-ANGLE HANDPIECE MUST BE CHECKED FOR EXTERNAL DAMAGE. BEFORE EACH USE, PERFORM A TEST RUN WITH THE CONTRA-ANGLE HANDPIECE, AND WATCH FOR ATYPICAL HEATING AND UNUSUAL NOISE AND VIBRATION. IMMEDIATELY STOP USING CONTRA-ANGLE HANDPIECES THAT ACT UNUSUAL. NEVER PRESS THE PUSHBUTTON DURING OPERATION. THIS ALSO INCLUDES LIFTING THE CHEEK OR TONGUE! TO ENSURE PROPER FUNCTION, THE MEDICAL DEVICE MUST BE SET UP ACCORDING TO THE REPROCESSING METHODS DESCRIBED IN THE KAVO INSTRUCTIONS FOR USE, AND THE CARE PRODUCTS AND CARE SYSTEMS DESCRIBED THEREIN MUST BE USED. KAVO RECOMMENDS SPECIFYING A SERVICE INTERVAL AT THE DENTAL OFFICE FOR A LICENSED SHOP TO CLEAN, SERVICE AND CHECK THE FUNCTIONING OF THE MEDICAL DEVICE. THIS SERVICE INTERVAL DEPENDS ON THE FREQUENCY OF USE AND SHOULD BE ADJUSTED ACCORDINGLY. (B)(4).
WHILE WORKING ON TOOTH# 30 THE HANDPIECE HEATED UP AND BURNED THE PATIENT ON CHEEK . THE DENTIST PRESCRIBED VICOPROFEN AND GAVE SOME VITAMINE E OINTMENT TO THE PATIENT TO APPLY TO BURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 545946 | GENTLEPOWER LUX CONTRA ANGLE 25LPR | ELECTRICAL DENTAL HANDPIECE | EKX | KAVO DENTAL GMBH | 25 LPR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Other |