CA090, DIRECT DRIVE LCA 3/BX
Report
- Report Number
- 2027111-2015-00509
- Event Type
- Malfunction
- Date Received
- August 14, 2015
- Date of Event
- July 21, 2015
- Report Date
- August 14, 2015
- Manufacturer
- APPLIED MEDICAL
- Product Code
- FZP
- PMA / PMN Number
- K011236
- Removal / Correction Number
- Z-1621-2016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
RA HAS JUST RECEIVED THE INCIDENT DEVICE AND HAS BEEN ASSIGNED TO ENGINEERING FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SENT UPON COMPLETION OF INVESTIGATION. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THE INITIAL REPORT WAS SUBMITTED, THEN THE SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA.
FULL UDI# PROVIDED. ON MARCH 18, 2016, APPLIED MEDICAL ISSUED A VOLUNTARY CLASS II RECALL OF THE CA090 DIRECT DRIVE CLIP APPLIER DUE TO INCREASED CUSTOMER FEEDBACK INDICATING INCONSISTENT CLIP APPLICATION, WHICH HAS THE POTENTIAL TO LEAD TO UNOCCLUDED VESSELS. AN INTERNAL CORRECTIVE ACTION REQUEST HAS BEEN INITIATED TO CONDUCT A THOROUGH INVESTIGATION AND IMPLEMENT APPROPRIATE CORRECTIVE ACTIONS. FDA HAS ISSUED RECALL NUMBER Z-1621-2016 FOR THIS RECALL. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THIS REPORT WAS SUBMITTED, THEN A SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA.
"WHEN PLACING THE CLIPS, THE CLIPS GET CROSSED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 540415 | CA090, DIRECT DRIVE LCA 3/BX | FZP | FZP | APPLIED MEDICAL | CA090 | 1240611 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |