1823260-2015-03973
Report
- Report Number
- 1823260-2015-03973
- Event Type
- Malfunction
- Date Received
- August 14, 2015
- Date of Event
- July 27, 2015
- Report Date
- September 2, 2015
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
THIS EVENT OCCURRED IN (B)(6). (B)(4).
THE TROPONIN T HS REAGENT LOT NUMBER WAS 181799. THE FIELD SERVICE ENGINEER INDICATED THE MAINTENANCE STATUS OF THE ANALYZER WAS ACCEPTABLE. A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE AVAILABLE DATA, A GENERAL REAGENT OR INSTRUMENT ISSUE IS NOT LIKELY. A POSSIBLE ROOT CAUSE MAY BE RELATED TO PRE-ANALYTICAL ISSUES. A POSSIBLE ELECTROMAGNETIC INTERFERENCE CANNOT BE EXCLUDED AS THERE ARE 2 CENTRIFUGES PLACED NEAR THE ANALYZER.
THE CUSTOMER COMPLAINED OF AN ERRONEOUS HIGH RESULT FOR 1 PATIENT SAMPLE TESTED FOR TROPONIN T HS (HIGH SENSITIVE) TROPONIN T HS BETWEEN 2 E601 ANALYZERS. THE ERRONEOUS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE INITIAL TROPONIN T HS RESULT ON THE FIRST E601 ANALYZER WAS 47.69 PG/ML. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. ON (B)(6) 2015 REPEAT TESTING WAS PERFORMED ON A 2ND E601 ANALYZER AND THE RESULTS WERE 9.95 PG/ML AND 9.97 PG/ML. THE PATIENT IS CURRENTLY HOSPITALIZED IN THE FIRST AID UNIT. NO ADVERSE EVENT DUE TO THE DEVICE HAS BEEN REPORTED. THE TROPONIN T HS REAGENT LOT NUMBER AND EXPIRATION DATE WERE NOT PROVIDED. THE FIELD SERVICE ENGINEER CHECKED THE EXTRA WASH SYSTEM OF THE ANALYZER AND FOUND A SMALL "BALL" OF DUST.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 079 YR |