OLYMPUS
Report
- Report Number
- 9611174-2015-00006
- Event Type
- Malfunction
- Date Received
- August 14, 2015
- Date of Event
- July 22, 2015
- Report Date
- July 22, 2015
- Manufacturer
- KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT) LTD
- Product Code
- FEM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
KEYMED LTD HAVE REQUESTED THE RETURN OF THE SUBJECT DEVICE FROM (B)(6), SO THAT A FULL INVESTIGATION CAN TAKE PLACE TO ESTABLISH THE ROOT CAUSE OF THIS MALFUNCTION. AWAITING RETURN OF SUBJECT DEVICE.
KEYMED LTD HAVE RECEIVED THE SUBJECT DEVICE. THE SUBJECT DEVICE HAS BEEN EVALUATED. THE EVALUATION DETERMINED THAT ONE SOCKET OUTLET ON THE CENTRAL STRIP WAS SEVERELY ERODED AROUND THE LEFT HAND POWER CONTACT DUE TO EXTREME HEAT. INTERNALLY THERE WAS SOME DISTORTION OF THE MOULDING WHERE THE CONTACT IS SECURED, BUT NOT TO THE SAME EXTENT AS EXTERNALLY. THE EQUIPMENT LEAD PLUG HAS NOT BEEN FULLY ENGAGED WITH THE SOCKET, SO THAT ITS LEFT HAND PIN WAS ONLY TOUCHING THE SOCKET CONTACT AT ITS TIP RATHER THAN BEING GRIPPED ON BOTH SIDES AS IT WOULD NORMALLY BE WHEN THE PLUG IS FULLY ENGAGED. THE SINGLE POINT OF CONTACT WILL PRODUCE CONSIDERABLE HEAT WHEN CURRENT IS FLOWING. THE IFU ALREADY CONTAINS CAUTIONS TO ENSURE THAT CONNECTORS ARE FULLY ENGAGED WHEN THE SYSTEM IS CONFIGURED AND TO CHECK THEM DURING ROUTINE MAINTENANCE.
THE OLYMPUS WM-NP2 MOBILE WORKSTATION IS INTENDED FOR USE IN MEDICAL FACILITIES UNDER THE DIRECTION OF A TRAINED PHYSICIAN, AND HAS BEEN DESIGNED TO BE USED WITH A RANGE OF OLYMPUS EQUIPMENT TO FACILITATE GI ENDOSCOPY, ENDOSCOPIC ULTRASOUND, RESPIRATORY AND SURGICAL ENDOSCOPIC PROCEDURES. THE SUBJECT DEVICE, A SEPARATION TRANSFORMER (B)(4), IS LOCATED ON THE UNDERSIDE OF THE BASE SHELF. THE SEPARATION TRANSFORMER CONNECTS TO THE MAINS POWER SUPPLY AND HAS A BANK OF OUTPUT SOCKETS WHICH ALLOWS FOR THE ELECTRICAL CONNECTION OF OLYMPUS MEDICAL DEVICES WHICH ARE SITUATED ON THE WORKSTATION. KEYMED (MEDICAL &INDUSTRIAL EQUIPMENT) LTD, HAS BEEN MADE AWARE OF AN EVENT IN (B)(6) WHERE DURING THE PREPARATION FOR USE, A SNAPPING NOISE OCCURRED AS THE FACILITY STAFF POWERED ON A (B)(4) LIGHT SOURCE WHICH WAS INSTALLED ON THE WORKSATATION. FACILITY STAFF CHECKED THE (B)(4) SEPARATION TRANSFORMER AND FOUND THAT THE OUTPUT SOCKET OF THE TRANSFORMER WHICH CONNECTED TO THE (B)(4) LIGHT SOURCE HAD MELTED. THE HEALTH FACILITY USED ANOTHER WORKSTATION TO PERFORM THE PROCEDURE. THERE IS NO REPORT OF INJURY TO PATIENT OR USER AND THEREFORE THIS REPORT IS SUBMITTED IN AN ABUNDANCE OF CAUTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 540454 | OLYMPUS | MAJ-1646 SEPARATION TRANSFORMER FOR THE WM-NP2 WORKSTATION | FEM | KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT) LTD | MAJ-1646 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |