FDA Adverse Event Injury Summary report: N

SUPERDIMENSION INREACH BRONCHUS SYSTEM

MDR report key: 5003073 · Received August 14, 2015

Report

Report Number
3004962788-2015-00058
Event Type
Injury
Date Received
August 14, 2015
Date of Event
July 21, 2015
Report Date
September 17, 2015
Manufacturer
SUPERDIMENSION INC.
Product Code
JAK
PMA / PMN Number
K092365
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SUPERDIMENSION HAS REQUESTED THE DEVICE FOR EVALUATION BUT HAS NOT RECEIVED ANYTHING TO DATE. THERE WERE NO ANOMALIES IDENTIFIED DURING THE INTERNAL REVIEW OF THE DHR OF THE SYSTEM CONSOLE. PNEUMOTHORAX IS A KNOWN SHORT TERM COMPLICATION WHEN A LUNG BIOPSY IS PERFORMED DURING A TRANSBRONCHIAL LUNG BIOPSY OR CT GUIDED PERCUTANEOUS BIOPSY. IF ADDITIONAL INFORMATION IS RECEIVED A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DATE OF FOLLOW-UP REPORT: 09/17/2015. A COMPONENT OF THE SUPERDIMENSION SYSTEM, CASE RECORDINGS, WERE RETURNED AND EVALUATED. THE EVALUATION RESULTED IN NO PROBLEM FOUND WITH THE SYSTEM. THE RECORDINGS CONFIRMED THE PHYSICIAN NAVIGATED BEYOND THE TARGET AND BREACHED THE PLEURA. THE PHYSICIAN STATED THAT HE EXPERIENCED NO DIFFICULTY WITH THE SUPERD NAVIGATION SYSTEM.

Description of Event or Problem · 1

THE PATIENT SUFFERED A PNEUMOTHORAX DURING A SUPERDIMENSION PROCEDURE AND RECEIVED A CHEST TUBE. IF ADDITIONAL INFORMATION PERTINENT TO THE INCIDENT IS OBTAINED A FOLLOW-UP REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
538314 SUPERDIMENSION INREACH BRONCHUS SYSTEM ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (COMMON NAME) JAK SUPERDIMENSION INC. AAS00161-20

Patients

Seq Age Sex Outcome Treatment
1 72 YR Other| R