VENTR CATH PVC 2 POS LST 18 FR

MDR report key: 5001390 · Received August 8, 2015

Report

Report Number
1718850-2015-00302
Event Type
Other
Date Received
August 8, 2015
Report Date
July 10, 2015
Manufacturer
SORIN GROUP USA, INC.
Product Code
DWF
PMA / PMN Number
K943934
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

NO SPECIFIC EVENT DATE WAS REPORTED. ADD'L DEVICE MFG DATES: 02/02/2015 & 02/16/2015. SORIN GROUP RECEIVED A REPORT THAT THE STYLET WAS VERY DIFFICULT TO EXTRACT. NO SPECIFIC EVENT DATE(S) WERE REPORTED. THE CUSTOMER PROVIDED THREE LOT NUMBER IN THIS ONE REPORT: 1502900021; EXP. DATE. 01/31/2015 & 1504300034 EXP DATE 02/28/2018. HOWEVER, NO SPECIFIC DETAILS WERE PROVIDED REGARDING THESE LOTS. NO UNIT HAS BEEN RETAINED FOR INVESTIGATION. THERE WAS NO REPORT OF PATIENT INJURY.

Additional Manufacturer Narrative · 1

SORIN GROUP RECEIVED A REPORT THAT THE STYLET WAS DIFFICULT TO EXTRACT. THERE WAS NO REPORT OF PATIENT INJURY. NO PRODUCT WAS RETURNED FOR EVALUATION. SORIN GROUP INITIATED A CAPA IN RESPONSE TO THIS ISSUE. EVALUATION OF UNITS FROM INVENTORY WAS ABLE TO REPRODUCE THE ISSUE OF DIFFICULTY REMOVING THE STYLET WHEN TESTED UNDER DRY CONDITIONS BUT ONCE FLUSHED, THE STYLET COULD BE REMOVED WITH NO DIFFICULTY. THE SORIN VENT CATHETER INSTRUCTIONS FOR USE INSTRUCT THE USER TO "REMOVE THE STYLET FROM THE CATHETER AND FLUSH THE CATHETER" PRIOR TO USE. THE CAPA INVESTIGATION LED TO THE CONCLUSION THAT THE ISSUE WAS CAUSED BY A PROBLEM WITH THE SURFACE FINISH ON THE STYLET. A REVIEW OF THE TUBING SPECIFICATION REVEALED THAT THE SURFACE FINISH DESCRIPTION WAS NOT SPECIFIC ENOUGH TO ENSURE THE FUNCTIONAL REQUIREMENTS WERE MET. THEREFORE THE SPECIFICATION WAS UPDATED TO CALL OUT THE SPECIFIC SURFACE FINISH STANDARD REQUIRED AND THIS UPDATE WAS COMMUNICATED TO THE SUPPLIER OF THE TUBING TO ENSURE PROPER IMPLEMENTATION. NO FURTHER ACTIONS ARE DEEMED NECESSARY AT THIS TIME.

Description of Event or Problem · 1

SORIN GROUP RECEIVED A REPORT THAT THE STYLET WAS VERY DIFFICULT TO EXTRACT. NO SPECIFIC EVENT DATE WAS REPORTED. THE CUSTOMER PROVIDED THREE LOT NUMBERS IN THIS ONE REPORT. HOWEVER, NO SPECIFIC DETAILS WERE PROVIDED REGARDING THESE LOTS. NO UNIT HAS BEEN RETAINED FOR INVESTIGATION. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
521178 VENTR CATH PVC 2 POS LST 18 FR CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS DWF SORIN GROUP USA, INC. NA 1502600129

Patients

Seq Age Sex Outcome Treatment
1 NP