FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4999097 · Received August 6, 2015

Report

Report Number
1052693-2015-01366
Event Type
Malfunction
Date Received
August 6, 2015
Date of Event
July 13, 2015
Report Date
December 9, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT NOT YET EVALUATED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER'S TEST STRIP HAD POOR STORAGE INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. ADDITIONAL PRODUCT CODES ADDED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMERWIFE STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMEREXPECTED BLOOD RESULTS ARE 95-105MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 02/28/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED IN THE KITCHEN AND WAS FIRST OPENED (B)(6) 2015. 1: 146MG/DL, (B)(6) 2015, 12:13PM, FASTING: YES; 2. 85MG/DL, (B)(6) 2015, 03:56AM, FASTING: YES; 3: 84MG/DL, (B)(6) 2015, 04:10AM, FASTING: YES; 4: 59MG/DL, (B)(6) 2015, 04:53AM, FASTING: YES; 5: 55MG/DL, (B)(6) 2015, 04:53AM, FASTING: YES. CUSTOMER'S CONCERN: 55MG/DL, (B)(6) 2015, 04:53AM, FASTING: YES. NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER'S WIFE STATES THAT HER FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 95-105MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 02/28/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED IN THE KITCHEN AND WAS FIRST OPENED (B)(6) 2015. (B)(6). FASTING:YES. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
520411 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2231

Patients

Seq Age Sex Outcome Treatment
1 0 YR