TRUERESULT
Report
- Report Number
- 1052693-2015-01366
- Event Type
- Malfunction
- Date Received
- August 6, 2015
- Date of Event
- July 13, 2015
- Report Date
- December 9, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT NOT YET EVALUATED. (B)(4).
(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER'S TEST STRIP HAD POOR STORAGE INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. ADDITIONAL PRODUCT CODES ADDED.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMERWIFE STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMEREXPECTED BLOOD RESULTS ARE 95-105MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 02/28/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED IN THE KITCHEN AND WAS FIRST OPENED (B)(6) 2015. 1: 146MG/DL, (B)(6) 2015, 12:13PM, FASTING: YES; 2. 85MG/DL, (B)(6) 2015, 03:56AM, FASTING: YES; 3: 84MG/DL, (B)(6) 2015, 04:10AM, FASTING: YES; 4: 59MG/DL, (B)(6) 2015, 04:53AM, FASTING: YES; 5: 55MG/DL, (B)(6) 2015, 04:53AM, FASTING: YES. CUSTOMER'S CONCERN: 55MG/DL, (B)(6) 2015, 04:53AM, FASTING: YES. NO ADVERSE EVENT REPORTED.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER'S WIFE STATES THAT HER FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 95-105MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 02/28/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED IN THE KITCHEN AND WAS FIRST OPENED (B)(6) 2015. (B)(6). FASTING:YES. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 520411 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2231 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |