FDA Adverse Event Malfunction Summary report: N

SPACELABS XPREZZON BEDSIDE MONITOR

MDR report key: 4996393 · Received August 12, 2015

Report

Report Number
3010157426-2015-00157
Event Type
Malfunction
Date Received
August 12, 2015
Date of Event
July 15, 2015
Report Date
September 23, 2015
Manufacturer
SPACELABS HEALTHCARE INC.
Product Code
MHX
PMA / PMN Number
K112962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

SPACELABS HAS LAUNCHED AN INVESTIGATION OF THIS ISSUE AND WILL FILE A SUPPLEMENTAL REPORT WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RECEIVED AT SPACELABS¿ EQUIPMENT SERVICE CENTER FOR REPAIR. THE REPORTED PROBLEM WAS VERIFIED. THE CPU PCBA 670-1488-00 WAS NOT FUNCTIONING; THEREFORE, THE PCBA WAS REPLACED. THE REPAIRED UNIT PASSED ALL FUNCTIONAL TESTS AND WAS RETURNED TO THE CUSTOMER. LACK OF POWER UP WAS THE WARNING SIGN THAT THE CUSTOMER NOTICED AND THAT PROMPTED THE ACTION OF REMOVING THE PRODUCT FROM USE. THIS REPORT IS COMPLETE AND THIS PARTICULAR ISSUE IS CONSIDERED CLOSED. SPACELABS MONITORS ITS COMPLAINTS FOR SIMILAR PRODUCT ISSUES TO DETERMINE NEED FOR ADDITIONAL INVESTIGATION.

Description of Event or Problem · 1

SPACELABS RECEIVED A REQUEST FOR SERVICE REPAIR FOR AN XPREZZON BEDSIDE MONITOR MODEL 91393 THAT STOPPED WORKING WHILE IS USE AND WOULD NOT RESTART ON (B)(6) 2015. NO ONE WAS INJURED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
530085 SPACELABS XPREZZON BEDSIDE MONITOR S-CLASS MULTIPARAMETER PATIENT MONITOR MHX SPACELABS HEALTHCARE INC. 91393

Patients

Seq Age Sex Outcome Treatment
1