FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM

MDR report key: 4993597 · Received August 11, 2015

Report

Report Number
3004753838-2015-80335
Event Type
Malfunction
Date Received
August 11, 2015
Date of Event
July 10, 2015
Report Date
July 13, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DOWNLOADED RECEIVER LOG CONFIRMED THE REPORTED EVENT OF THE RECEIVER DISPLAYING THE INITIALIZING SCREEN WITHOUT A MANUAL RESTART. A ROOT CAUSE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015, TO REPORT THAT ON (B)(6) 2015, THEIR RECEIVER DISPLAYED THE INITIALIZING SCREEN WITHOUT MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
523445 DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM MDS MDS DEXCOM, INC. MT22608-BLU 5197967

Patients

Seq Age Sex Outcome Treatment
1 10 YR