FDA Adverse Event Malfunction Summary report: N

E SERIES DEFIBRILLATOR

MDR report key: 4993311 · Received August 6, 2015

Report

Report Number
1220908-2015-01992
Event Type
Malfunction
Date Received
August 6, 2015
Report Date
July 17, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946015640
PMA / PMN Number
K042007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY FLEX CABLE. THE FLEX CABLE WAS REPLACED TO REMEDY THE PROBLEM. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE WAS INTERMITTENTLY UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
520053 E SERIES DEFIBRILLATOR E SERIES MKJ ZOLL MEDICAL CORPORATION E SERIES NA 00847946015640

Patients

Seq Age Sex Outcome Treatment
1 NA