FDA Adverse Event
Malfunction
Summary report: N
E SERIES DEFIBRILLATOR
MDR report key: 4993311
·
Received August 6, 2015
Report
- Report Number
- 1220908-2015-01992
- Event Type
- Malfunction
- Date Received
- August 6, 2015
- Report Date
- July 17, 2015
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- UDI-DI
- 00847946015640
- PMA / PMN Number
- K042007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY FLEX CABLE. THE FLEX CABLE WAS REPLACED TO REMEDY THE PROBLEM. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE WAS INTERMITTENTLY UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 520053 | E SERIES DEFIBRILLATOR | E SERIES | MKJ | ZOLL MEDICAL CORPORATION | E SERIES | NA | 00847946015640 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |