FDA Adverse Event Malfunction Summary report: N

LIGACLIP CLIP APPLIER

MDR report key: 4992993 · Received August 11, 2015

Report

Report Number
3005075853-2015-05082
Event Type
Malfunction
Date Received
August 11, 2015
Date of Event
July 23, 2015
Report Date
July 28, 2015
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BATCH # M9117G. THE ER320 DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT THE DEVICE WAS FUNCTIONALLY EVALUATED. UPON FIRING OF THE DEVICE, THE REMAINING CLIPS WERE EJECTED DUE TO THE CONDITION OF THE JAWS; FINALLY THE INSTRUMENT LOCKED OUT AS INTENDED. POSSIBLE CAUSES FOR THE FOUND CONDITION OF THE YIELDED JAWS MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED BY THE AFFILIATE THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS FROM THE APPLIER FELL DOWN. CROSS FIRING HAPPENED ON THE CLIPS THAT WERE USED FOR LIGATION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
526109 LIGACLIP CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1