FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 4992829 · Received August 7, 2015

Report

Report Number
9615739-2015-00244
Event Type
Malfunction
Date Received
August 7, 2015
Date of Event
July 16, 2015
Report Date
July 16, 2015
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CASTERS AND SUPPORTS WERE BROKEN AT THE SHAFT. PER THE HILL-ROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. CHECK THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE PEDALS ARE PRESSED AND REPAIR AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED FROM 2012 2015. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE CASTERS AND SUPPORTS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
521258 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1