FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4991097 · Received August 11, 2015

Report

Report Number
2032227-2015-26596
Event Type
Malfunction
Date Received
August 11, 2015
Date of Event
July 15, 2015
Report Date
July 16, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT RESPONSE FROM ESC BUTTON, DUE TO CORRODED KEYPAD TRACES. THE DEVICE WAS ALSO RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE LCD WINDOW, CRACKED LCD WINDOW, AND A SCRATCHED RESERVOIR TUBE WINDOW. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CALLED AND REPORTED THAT NONE OF THE BUTTONS ON THE INSULIN PUMP WERE WORKING. CUSTOMER'S BLOOD GLUCOSE WAS 232 MG/DL. CUSTOMER WAS ADVISED THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
525338 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754CAS