CA090, DIRECT DRIVE LCA 3/BX
Report
- Report Number
- 2027111-2015-00545
- Event Type
- Malfunction
- Date Received
- August 10, 2015
- Date of Event
- April 15, 2015
- Report Date
- August 10, 2015
- Manufacturer
- APPLIED MEDICAL
- Product Code
- FZP
- PMA / PMN Number
- K011236
- Removal / Correction Number
- Z-1621-2016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
FULL UDI# PROVIDED. ON MARCH 18, 2016, APPLIED MEDICAL ISSUED A VOLUNTARY CLASS II RECALL OF THE CA090 DIRECT DRIVE CLIP APPLIER DUE TO INCREASED CUSTOMER FEEDBACK INDICATING INCONSISTENT CLIP APPLICATION, WHICH HAS THE POTENTIAL TO LEAD TO UNOCCLUDED VESSELS. AN INTERNAL CORRECTIVE ACTION REQUEST HAS BEEN INITIATED TO CONDUCT A THOROUGH INVESTIGATION AND IMPLEMENT APPROPRIATE CORRECTIVE ACTIONS. FDA HAS ISSUED RECALL NUMBER Z-1621-2016 FOR THIS RECALL. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THIS REPORT WAS SUBMITTED, THEN A SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA. THIS REPORT IS TO FOLLOW UP WITH MEDWATCH REPORT# MW5042179.
THE INCIDENT DEVICE ANTICIPATED TO RETURN. A FOLLOW-UP REPORT WILL BE PROVIDED UPON COMPLETION OF INVESTIGATION. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THE INITIAL REPORT WAS SUBMITTED, THEN THE SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA.
LAPAROSCOPIC CHOLECYSTECTOMY- "APPLIED MEDICAL DIRECT DRIVE, DISPOSABLE LAPAROSCOPIC CLIP APPLIER WAS OPENED, NEW, ONTO THE SURGICAL FIELD AND ONLY CONTAINED (1) CLIP. THIS WAS EXCHANGE FOR A "LIKE" DEVICE WHICH HAD THE (20) INDICATED CLIPS. DIAGNOSIS OR REASON FOR USE: LAPAROSCOPIC CHOLECYSTECTOMY." PATIENT STATUS- "STABLE AT THE COMPLETION OF SURGERY."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 523125 | CA090, DIRECT DRIVE LCA 3/BX | FZP | FZP | APPLIED MEDICAL | CA090 | 1240610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |