FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4988749 · Received August 10, 2015

Report

Report Number
3004753838-2015-80432
Event Type
Malfunction
Date Received
August 10, 2015
Date of Event
July 16, 2015
Report Date
July 16, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5201173), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(4) 2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED A PERMANENT OUT OF RANGE SIGNAL. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
523130 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649 5179428

Patients

Seq Age Sex Outcome Treatment
1 72 YR