DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM, DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2015-80348
- Event Type
- Malfunction
- Date Received
- August 10, 2015
- Date of Event
- July 13, 2015
- Report Date
- July 13, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE TRANSMITTER (B)(4) THAT WAS USED WITH THE DEVICE WAS RETURNED FOR EVALUATION ON (B)(6) 2015. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE RETURNED TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5201166), WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED A PERMANENT OUT OF RANGE SIGNAL. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 522782 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM, DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22608-BLU | 5201275 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |