INNER SHEATH, FOR 26 FR. OUTER SHEATH
Report
- Report Number
- 9610773-2015-00040
- Event Type
- Malfunction
- Date Received
- August 10, 2015
- Date of Event
- July 29, 2015
- Report Date
- July 29, 2015
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- FJL
- PMA / PMN Number
- K931995
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC), (B)(4)(RETURNED TO OMSC ON 2015-07-31). THE EVALUATION/INVESTIGATION CONFIRMED THE REPORTED PHENOMENON OF A BROKEN/FRACTURED CERAMIC INSULATION AT THE DISTAL END OF THE INNER SHEATH. A LARGE CERAMIC FRAGMENT IS BROKEN OFF. HOWEVER, NO PARTS ARE ACTUALLY MISSING. CAUSAL FOR THIS DAMAGE AND BREAKAGE OF THE CERAMIC INSULATION IS MECHANICAL OVERLOAD BY THE APPLICATION OF EXCESSIVE FORCE LIKE IMPACT, FALL, SHOCK OR SIMILAR STRESS. IT IS CLEARLY STATED AS A WARNING NOTE IN THE INSTRUCTIONS THAT IMPACT, FALL, SHOCK OR SIMILAR STRESS CAN DAMAGE THE CERAMIC INSULATION AT THE SHEATH'S DISTAL END AND THAT THE INSTRUMENT MUST NOT BE USED IF DAMAGED AS OTHERWISE THIS CAN CAUSE INJURIES TO THE PATIENT AND/OR USER. THE USER APPARENTLY DID NOT FOLLOW THESE INSTRUCTIONS AS THE DAMAGE AND BREAKAGE OF THE CERAMIC INSULATION WERE CAUSED BY MECHANICAL OVERLOAD. THEREFORE THIS EVENT/INCIDENT WAS ATTRIBUTED TO ABNORMAL USE/OFF-LABEL USE AND THE CASE WILL BE CLOSED FROM OLYMPUS SIDE WITH NO FURTHER ACTIONS. HOWEVER, THE EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES AND THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS AND PRO RE NATA TRAINED AGAIN ON THE CORRECT USAGE OF THE OLYMPUS MEDICAL DEVICES. OLYMPUS SUBMITS THIS EVENT/INCIDENT AS A MEDICAL DEVICE REPORT (MDR) IN ABUNDANCE OF CAUTION.
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE BLADDER TUMOR (TURBT) PROCEDURE, THE SURGICAL TEAM NOTICED THAT THE CERAMIC INSULATION AT THE DISTAL END OF THE INNER SHEATH HAD BROKEN/FRACTURED. IT IS UNKNOWN WHEN EXACTLY THIS DAMAGE OCCURRED AND WHETHER A FRAGMENT/PART FELL INSIDE THE PATIENT. HOWEVER, IT WAS CONFIRMED THAT NO FRAGMENTS/PARTS REMAINED INSIDE THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED BUT THE INTENDED PROCEDURE WAS REPORTEDLY COMPLETED WITH THE SAME INNER SHEATH AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY. IN ADDITION, IT WAS REPORTED THAT THE INNER SHEATH HAD BEEN INSPECTED BEFORE THE PROCEDURE AND THAT NO ABNORMALITIES OR IRREGULARITIES WERE DETECTED AT THAT TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 522184 | INNER SHEATH, FOR 26 FR. OUTER SHEATH | RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS | FJL | OLYMPUS WINTER & IBE GMBH | 10XW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |