FDA Adverse Event Injury Summary report: N

LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI)

MDR report key: 4975933 · Received August 4, 2015

Report

Report Number
2023950-2015-00121
Event Type
Injury
Date Received
August 4, 2015
Date of Event
June 24, 2015
Report Date
August 4, 2015
Manufacturer
ZEST ANCHORS, INC.
Product Code
DZE
PMA / PMN Number
K120198
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

USER DOCUMENTATION (TECHNIQUE MANUAL, P/N L8019-TM) SPECIFIES THAT IMMEDIATE LOADING IS SUITABLE ONLY IF SUFFICIENT PRIMARY STABILITY OF THE IMPLANT IS ACHIEVED AT THE TIME OF PLACEMENT. ALSO, THE RECOMMENDED IMPLANT INSERTION TORQUE AND IMPLANT ATTACHMENT TORQUE IS 30 N-CM. IF THE IMPLANT PLACEMENT TORQUE IS LESS THAN 30 N-CM, THEN THE ATTACHMENTS SHOULD ONLY BE HAND TIGHTENED. THE IMPLANT INSERTION TORQUE SHOULD NOT EXCEED 70 N-CM; IF 70 N-CM IS REACHED PRIOR TO FULL SEATING, THE IMPLANT SHOULD BE REMOVED AND THE OSTEOTOMY SHOULD BE ENLARGED. CLINICIAN STATED THAT LODI IMPLANT FAILED TO OSSEOINTEGRATE. PATIENT HAS LOW DENSITY BONE. THE IMPLANT WAS NOT IMMEDIATELY LOADED. THE CLINICIAN DID NOT PROVIDE THE FOLLOWING INFORMATION: AMOUNT OF TORQUE USED ON ABUTMENT & IMPLANT; WHETHER PRIMARY STABILITY WAS ACHIEVED. FAILURE TO OSSEOINTEGRATE IS A WELL-DOCUMENTED INHERENT RISK OF DENTAL IMPLANTS. IN THE MAJORITY OF CASES WHERE AN IMPLANT FAILS TO INTEGRATE WITH THE BONE AND IS REJECTED BY THE BODY, THE CAUSE IS UNK. SINCE THE DISTRIBUTOR WAS UNABLE TO PROVIDE THE LOT NUMBER, ZEST WAS UNABLE TO REVIEW THE SPECIFIC LHR RECORDS. HOWEVER, ALL ZEST PRODUCTS THAT ARE SHIPPED OUT MUST MEET THEIR SPECIFIC PRODUCT SPECS AND MUST BE APPROVED FOR PRODUCT RELEASE. ANY ISSUES ARE SUCCESSFULLY RESOLVED PRIOR TO THE RELEASE OF ALL PARTS. NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

LODI IMPLANTS FAILED TO OSSEOINTEGRATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509576 LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI) ENDOSSEOUS DENTAL IMPLANT DZE ZEST ANCHORS, INC. NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention