SMALL BATTERY DRIVE
Report
- Report Number
- 3009450871-2015-12678
- Event Type
- Malfunction
- Date Received
- August 5, 2015
- Date of Event
- June 22, 2015
- Report Date
- July 11, 2015
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE LOT NUMBER WAS DOCUMENTED AS 2427 IN THE INITIAL REPORT. THE DEVICE LOT NUMBER HAS BEEN UPDATED AS 003454. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE WAS FUNCTIONAL. THEREFORE, THE REPORTED CONDITION WAS NOT CONFIRMED. AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. HOWEVER, DURING EVALUATION, IT WAS OBSERVED THAT THE TRIGGER WAS JAMMED AND ITS COMPONENTS WERE WORN. IT WAS DETERMINED THAT THIS WAS DUE TO NORMAL USE OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE/TESTING, IT WAS OBSERVED THAT THE SMALL BATTERY DRIVE DEVICE WOULD NOT SPIN THE DRILL BIT. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 517025 | SMALL BATTERY DRIVE | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | DEPUY SYNTHES POWER TOOLS | 003454 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |