FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4974678 · Received August 5, 2015

Report

Report Number
2531779-2015-27107
Event Type
Malfunction
Date Received
August 5, 2015
Report Date
July 21, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 09/29/2015-PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/10/2015 WITH THE FOLLOWING FINDINGS: THE PUMP FAILED TO POWER ON. MOISTURE WAS OBSERVED BEHIND THE PUMP¿S DISPLAY LENS. A PUMP CASE CRACK WAS OBSERVED. LEAK TESTING REVEALED A PUMP CASE LEAK. MOISTURE WAS OBSERVED ON THE PUMP¿S INTERNAL COMPONENTS.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A NO POWER (MOISTURE INGRESS-BEHIND DISPLAY) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
516512 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION