EON MINI IPG, 16-CHANNEL RECHARGEABLE
Report
- Report Number
- 1627487-2015-20431
- Event Type
- Injury
- Date Received
- August 5, 2015
- Date of Event
- June 25, 2015
- Report Date
- July 31, 2015
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-05242011-002-R
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). FIELD ADVISORY #'S: 1627487-07262012-002-R; 1627487-12192011-003-R. THIS IPG SERIAL NUMBER WAS INCLUDED IN FIELD ADVISORIES. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY. (B)(4).
SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
IT WAS REPORTED THE PATIENT WAS UNABLE TO ESTABLISH COMMUNICATION BETWEEN HER IPG AND MULTIPLE EXTERNAL DEVICES AFTER NOT USING/CHARGING THE DEVICE. A SJM REPRESENTATIVE CONFIRMED THE COMMUNICATION ISSUE. SURGICAL INTERVENTION WILL BE TAKEN AT A LATER DATE TO ADDRESS THE ISSUE.
FURTHER FOLLOW UP IDENTIFIED THE IPG WAS EXPLANTED AND REPLACED WHICH RESOLVED THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 515062 | EON MINI IPG, 16-CHANNEL RECHARGEABLE | SCS IPG | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 2846379 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Other | MODEL 1194(2), SCS ANCHOR |