FDA Adverse Event Malfunction Summary report: N

UNKNOWN_ENDOSCOPY_PRODUCT

MDR report key: 4974155 · Received August 5, 2015

Report

Report Number
0002936485-2015-00708
Event Type
Malfunction
Date Received
August 5, 2015
Date of Event
July 9, 2015
Report Date
July 9, 2015
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
BZN
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE REPORTED DEVICE WAS NOT RECEIVED FOR INVESTIGATION; THEREFORE THE REPORTED FAILURE CANNOT BE CONFIRMED. THE COMPLAINT WILL BE CLOSED WITHOUT A DETAILED INVESTIGATION REPORT AND BASED ON PROBABLE ROOT CAUSE. IN THE EVENT THAT THE DEVICE IS RECEIVED, THE COMPLAINT WILL BE REOPENED, A FULL EVALUATION WILL BE CONDUCTED, AND THE INVESTIGATION WILL BE UPDATED WITH THE NEW RESULTS. PROBABLE ROOT CAUSE FOR THE REPORTED FAILURE INVOLVING THIS DEVICE COULD HAVE BEEN CAUSED BY: TOO MUCH HEAT. TRANSFORMER DESIGN. CART DESIGN. MANUFACTURING NONCONFORMITY (TEKNA). USE ERROR. THE PRODUCT WAS NOT RETURNED FOR INVESTIGATION THEREFORE THE REPORTED FAILURE MODE WAS NOT CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE POWER CABLE FROM THE STRYKER TOWER SPARKED DURING THE CASE. SPARKING OCCURRED AT THE PLUG OUTLET AFTER TURNING THE LASER AND TOWER ON. THE ELECTRICIAN AND FACILITY ADMINISTRATOR BOTH ACKNOWLEDGE THAT THE STRYKER TOWER WAS NOT AT FAULT FOR THE ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE POWER CABLE FROM THE STRYKER TOWER SPARKED DURING THE CASE. SPARKING OCCURRED AT THE PLUG OUTLET AFTER TURNING THE LASER AND TOWER ON. THE ELECTRICIAN AND FACILITY ADMINISTRATOR BOTH ACKNOWLEDGE THAT THE STRYKER TOWER WAS NOT AT FAULT FOR THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
517799 UNKNOWN_ENDOSCOPY_PRODUCT CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT) BZN STRYKER ENDOSCOPY-SAN JOSE

Patients

Seq Age Sex Outcome Treatment
1