FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4973901 · Received August 5, 2015

Report

Report Number
2531779-2015-27053
Event Type
Malfunction
Date Received
August 5, 2015
Report Date
July 21, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 09/16/2015-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/27/2015 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX SHOWED MULTIPLE LOSS OF PRIME WARNINGS ASSOCIATED WITH LOW NON-ZERO FORCE. DURING TESTING, THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIME. THE FORCE SENSOR CALIBRATION WAS FOUND TO BE OUT OF SPECIFICATIONS. THE FORCE SENSOR RESISTANCE WAS TESTED AND WAS FOUND TO BE WITHIN SPECIFICATIONS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT THE PUMP REPEATEDLY EMITTED LOSS OF PRIME WARNINGS DESPITE CARTRIDGE CHANGES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
516293 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 24 YR