LIFEVEST WCD 4000 SYSTEM
MDR report key: 4973385
·
Received August 5, 2015
Report
- Report Number
- 3008642652-2015-04662
- Event Type
- Malfunction
- Date Received
- August 5, 2015
- Date of Event
- July 6, 2015
- Report Date
- July 27, 2015
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETS) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS RESETTING. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN Y1 CRYSTAL OSCILLATOR ON THE BEDSIDE PCA. THE ROOT CAUSE FOR THE OPEN CRYSTAL OSCILLATOR COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED BATTERY CHARGER/MODEM SN (B)(4) AND REPORTED THAT THE BATTERY CHARGER/MODEM WAS RESETTING.