SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2015-08344
- Event Type
- Malfunction
- Date Received
- August 5, 2015
- Date of Event
- July 1, 2015
- Report Date
- July 20, 2015
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- Z-1484-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. THE EVALUATION WAS ABLE TO CONFIRM THE SYSTEM ERROR 322 ALARM THROUGH REVIEW OF THE DEVICE EVENT HISTORY LOG. THE ALARM COULD NOT BE REPRODUCED AND A CAUSE COULD NOT BE IDENTIFIED. THE PUMP WAS TESTED WITH PASSING RESULTS. BAXTER HAS INITIATED AN APPROVED REMEDIAL ACTION; HOWEVER THIS DEVICE HAS A VERSION OF SOFTWARE THAT DOES NOT HAVE AN APPROVED SOFTWARE UPDATE. THE DEVICE WAS FOUND TO BE OPERATING MECHANICALLY AS DESIGNED AND NO ACTION WILL BE TAKEN AT THIS TIME. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE. BAXTER HAS INITIATED A CAPA TO FURTHER INVESTIGATE THIS ISSUE.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION. THE EVALUATION HAS NOT BEEN COMPLETED AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE PROVIDED WHEN THE INVESTIGATION IS COMPLETED.
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 516041 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |