FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 4972166 · Received August 5, 2015

Report

Report Number
2649622-2015-10274
Event Type
Injury
Date Received
August 5, 2015
Date of Event
June 1, 2015
Report Date
June 1, 2015
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT COMPLAINED OF POCKET PAIN AND DISCOMFORT AT THE ACCESS SITE DUE TO A SMALL SUBCLAVIAN VEIN. BOTH THE ATRIAL AND RIGHT VENTRICULAR LEADS WERE EXPLANTED AND REPLACED WITH A CEPHALIC ACCESS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
513124 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00052 YR Hospitalization| R A2DR01 IPG, 5086MRI LEAD