FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4972145 · Received August 5, 2015

Report

Report Number
2649622-2015-10280
Event Type
Injury
Date Received
August 5, 2015
Date of Event
May 11, 2015
Report Date
May 11, 2015
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INTERMITTENT PHRENIC NERVE STIMULATION. CHEST X-RAY CONFIRMED THAT THE RIGHT ATRIAL (RA) LEAD SLACK HAD PULLED BACK AND THE LEAD HAD DISLODGED. THE LEAD TIP WAS POSITIONED CLOSE THE SUPERIOR VENA CAVA (SVC)-RA JUNCTION. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514166 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Hospitalization| R 6935M62 LEAD, DTBA1D4 CRT-D