FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 4970657 · Received August 5, 2015

Report

Report Number
2183613-2015-00706
Event Type
Malfunction
Date Received
August 5, 2015
Date of Event
April 17, 2015
Report Date
May 8, 2015
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THE BATTERY CONTACTS WERE COMPRESSED AND THE BATTERY DRAWER WAS BROKEN. ANALYSIS ALSO FOUND THE UPPER CASE, LOWER CASE, TWO SIDE BAIL COVERS, AND THE RING COVER WERE BROKEN. TWO SIDE BAILS WERE MISSING AND THE KEYBOARD PAD WAS CONTAMINATED. (B)(4).

Description of Event or Problem · 1

THE EXTERNAL PULSE GENERATOR WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
518737 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1