FDA Adverse Event
Malfunction
Summary report: N
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
MDR report key: 4970657
·
Received August 5, 2015
Report
- Report Number
- 2183613-2015-00706
- Event Type
- Malfunction
- Date Received
- August 5, 2015
- Date of Event
- April 17, 2015
- Report Date
- May 8, 2015
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THE BATTERY CONTACTS WERE COMPRESSED AND THE BATTERY DRAWER WAS BROKEN. ANALYSIS ALSO FOUND THE UPPER CASE, LOWER CASE, TWO SIDE BAIL COVERS, AND THE RING COVER WERE BROKEN. TWO SIDE BAILS WERE MISSING AND THE KEYBOARD PAD WAS CONTAMINATED. (B)(4).
Description of Event or Problem · 1
THE EXTERNAL PULSE GENERATOR WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.