FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4969981 · Received August 5, 2015

Report

Report Number
3004753838-2015-60255
Event Type
Malfunction
Date Received
August 5, 2015
Date of Event
July 15, 2015
Report Date
July 15, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE WASN'T RETURNED, BUT THE RECEIVER THAT WAS USED WITH THE DEVICE HAS BEEN RECEIVED FOR EVALUATION ON (B)(6) 2015. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE RETURNED SHARE RECEIVER (PART NUMBER STR-DR-001/SERIAL NUMBER (B)(4)/LOT NUMBER 5200908), WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DIAGNOSTIC CHART CONFIRMED THE REPORTED EVENT OF AN INTERMITTENT OUT OF RANGE SIGNAL. A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT AN INTERMITTENT OUT OF RANGE SIGNAL ON (B)(6) 2015. AT THE TIME OF CONTACT THE PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
512344 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9438-05 5201333

Patients

Seq Age Sex Outcome Treatment
1 47 YR