FDA Adverse Event Malfunction Summary report: N

LIGACLIP CLIP APPLIER

MDR report key: 4965636 · Received August 4, 2015

Report

Report Number
3005075853-2015-04970
Event Type
Malfunction
Date Received
August 4, 2015
Date of Event
July 14, 2015
Report Date
July 20, 2015
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: BATCH # M9125V. CONCLUSION: THE ANALYSIS RESULTS OF THE ER320 DEVICE FOUND THAT IT WAS RECEIVED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE MISALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY; DURING THE ANALYSIS, THE INSTRUMENT WAS CYCLED AND IT FED AND FORMED FIFTEEN SCISSORED CLIPS DUE TO THE MISALIGNED CONDITION OF THE JAWS. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC COLON 'KIR' PROCEDURE, THE CLIPS COME OFF THE VESSELS WHEN THEY ARE TOO SECURE/GOING CROOKED ON A VESSEL. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509839 LIGACLIP CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA M4HG81

Patients

Seq Age Sex Outcome Treatment
1