FDA Adverse Event Injury Summary report: N

2.7MM/3.5MM VA-LCP LAT DISTAL HUMERUS PL 1H/LT/69MM-SHORT

MDR report key: 4964077 · Received August 4, 2015

Report

Report Number
2520274-2015-15110
Event Type
Injury
Date Received
August 4, 2015
Report Date
July 10, 2015
Manufacturer
SYNTHES (USA)
Product Code
HRS
PMA / PMN Number
PK120070
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) PROCEDURE TO TREAT A LEFT SUPRACONDYLAR HUMERUS FRACTURE WITH AND UNKNOWN NUMBER OF SCREWS AND TWO PLATES ON (B)(6) 2015. THE PATIENT WAS INVOLVED IN A MOTOR VEHICLE ACCIDENT ON UNKNOWN DATE AFTER IMPLANT SURGERY. THE PATIENT REPORTED BLISTERING AND A SCREW HEAD PROTRUDING OVER LATERAL ASPECT OF ELBOW FOLLOWING MOTOR VEHICLE ACCIDENT. THE PATIENT RETURNED TO OPERATING ROOM ON (B)(6) 2015 FOR REMOVAL OF THE PROTRUDING LOCKING SCREW. A WOUND CULTURE WAS ALSO TAKEN DURING THE REVISION SURGERY. ON (B)(6) 2015 THE CULTURE RESULTS WERE DETERMINED TO BE POSITIVE FOR INFECTION; HOWEVER, THE TYPE OF ORGANISM UNKNOWN. THE PATIENT DID NOT FOLLOW UP WITH RECOMMENDATION FOR ANTIBIOTIC THERAPY AND RETURNED TO SURGEON OFFICE ON (B)(6) 2015, WITH DRAINING PUS FROM WOUND. AN X-RAY TAKEN ON (B)(6) 2015 REVEALED A BROKEN LOCKING SCREW. THE PATIENT RETURNED TO THE OPERATING ON (B)(6) 2015 AND HAD LOCKING SCREW AND TWO PLATES REMOVED DUE TO BROKEN LOCKING SCREW, INFECTION AND LOSS OF REDUCTION. THE PATIENT WAS RE-PLATED SUCCESSFULLY. THERE WAS NO FURTHER PATIENT HARM OR ADDITIONAL INFORMATION REPORTED. THIS REPORT ADDRESSES THE FIRST REVISION SURGERY INVOLVING THE REMOVAL OF THE PROTRUDING LOCKING SCREW. THE SECOND REVISION SURGERY AND EVENTS ARE ADDRESSED AND REPORTED UNDER A SEPARATE COMPLAINT NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
508008 2.7MM/3.5MM VA-LCP LAT DISTAL HUMERUS PL 1H/LT/69MM-SHORT PLATE,FIXATION,BONE HRS SYNTHES (USA)

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention