AMISTEM C CEMENTED STD STEM #3
Report
- Report Number
- 3005180920-2015-00166
- Event Type
- Injury
- Date Received
- July 31, 2015
- Report Date
- November 4, 2015
- Manufacturer
- MEDACTA INTERNATIONAL, SA
- Product Code
- LZO
- PMA / PMN Number
- K103189
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- NOT APPLICABLE
Narratives
ON 05 OCTOBER 2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON 07 OCTOBER 2015 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
BATCH REVIEW PERFORMED ON JULY 31, 2015. LOT 140787: (B)(4) STEMS MANUFACTURED AND RELEASED ON APRIL 7, 2014. NO ANOMALIES FOUND. TO DATE, (B)(4) ITEMS OF THE LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR ISSUE REPORTED. CLINICAL EVAL PERFORMED ON JULY 24,2015 BY (B)(4). THE STEM APPEARS TO BE EMBEDDED IN AN UNEVEN CEMENT MANTLE. CEMENT IS REMARKABLY DEFICIENT ON ANTERIOR/PROXIMAL AND DISTAL PORTION OF THE STEM. IN THESE CONDITIONS, SINKING OF THE STEM IS A POSSIBLE COMPLICATION. WE HAVE NO INFO AS TO THE POSSIBLE CAUSE OF SUBOPTIMAL CEMENTATION, WHICH APPEARS TO BE THE ROOT CAUSE FOR REVISION. ON JULY 14, 2015, THE SALES REPORTED THAT THE PIECES WILL NOT AVAILABLE FOR FURTHER INVESTIGATION.
(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 498918 | AMISTEM C CEMENTED STD STEM #3 | CEMENTED FEMORAL STEM | LZO | MEDACTA INTERNATIONAL, SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |