FDA Adverse Event Injury Summary report: N

AMISTEM C CEMENTED STD STEM #3

MDR report key: 4963986 · Received July 31, 2015

Report

Report Number
3005180920-2015-00166
Event Type
Injury
Date Received
July 31, 2015
Report Date
November 4, 2015
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
LZO
PMA / PMN Number
K103189
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ON 05 OCTOBER 2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON 07 OCTOBER 2015 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON JULY 31, 2015. LOT 140787: (B)(4) STEMS MANUFACTURED AND RELEASED ON APRIL 7, 2014. NO ANOMALIES FOUND. TO DATE, (B)(4) ITEMS OF THE LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR ISSUE REPORTED. CLINICAL EVAL PERFORMED ON JULY 24,2015 BY (B)(4). THE STEM APPEARS TO BE EMBEDDED IN AN UNEVEN CEMENT MANTLE. CEMENT IS REMARKABLY DEFICIENT ON ANTERIOR/PROXIMAL AND DISTAL PORTION OF THE STEM. IN THESE CONDITIONS, SINKING OF THE STEM IS A POSSIBLE COMPLICATION. WE HAVE NO INFO AS TO THE POSSIBLE CAUSE OF SUBOPTIMAL CEMENTATION, WHICH APPEARS TO BE THE ROOT CAUSE FOR REVISION. ON JULY 14, 2015, THE SALES REPORTED THAT THE PIECES WILL NOT AVAILABLE FOR FURTHER INVESTIGATION.

Description of Event or Problem · 1

(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498918 AMISTEM C CEMENTED STD STEM #3 CEMENTED FEMORAL STEM LZO MEDACTA INTERNATIONAL, SA

Patients

Seq Age Sex Outcome Treatment
1